Abstract
Background
Cementless glass ceramics containing apatite and wollastonite (AW-GC) bottom-coated
titanium hip implants were developed; early excellent clinical and radiographic results
have been reported previously. This study aimed to investigate the long-term clinical
and radiographic outcomes in detail, and the wear rate of HXLPE.
Methods
We retrospectively evaluated 99 patients (117 hips) between November 2001 and December
2007. The survival rate was estimated using the Kaplan–Meier method. Hip joint function
was evaluated using the Japanese Orthopaedic Association (JOA) score. The extent of
radiographic signs was determined from the radiographs performed at the last follow-up.
Polyethylene wear was measured using Martell's Hip Analysis Suite. We assessed the
possible factors affecting the steady-state linear wear rate.
Results
The mean follow-up period was 14.8 ± 2.1 (10–18.6) years. The mean JOA score improved
to 88.7 ± 9.4 (59–100) at the final follow-up from 47.8 ± 12.5 (17–76) before surgery.
The overall survival rate with the end point of all-cause revision and wear-related
revision was 99% and 100% respectively. There was no osteolysis or loosening of either
the acetabular or femoral component. All hips were classified as having bone ingrowth
fixation. The mean steady-state wear rate was 0.008 ± 0.025 mm/year. We found no significant
correlation between the wear rate and age, body weight, body mass index, cup inclination
and femoral head size.
Conclusion
The combination of AW-GC bottom-coated implants and HXLPE showed excellent implant
survival and wear resistance for 15 years.
Keywords
To read this article in full you will need to make a payment
Purchase one-time access:
Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online accessOne-time access price info
- For academic or personal research use, select 'Academic and Personal'
- For corporate R&D use, select 'Corporate R&D Professionals'
Subscribe:
Subscribe to Journal of Orthopaedic ScienceAlready a print subscriber? Claim online access
Already an online subscriber? Sign in
Register: Create an account
Institutional Access: Sign in to ScienceDirect
References
- Fatigue and life-time of bioactive glass-ceramic A-W containing apatite and wollastonite.J Mater Sci. 1987 Nov 1; 22: 4067-4070
- Cementless total hip replacement. Bio-active glass ceramic coating studied in dogs.Acta Orthop Scand. 1993 Dec; 64: 607-612
- Good short-term outcome of primary total hip arthroplasty with cementless bioactive glass ceramic bottom-coated implants: 109 hips followed for 3–9 years.Acta Orthop. 2012 Dec; 83: 599-603
- Highly cross-linked polyethylene in total hip arthroplasty in patients younger than 50 Years with osteonecrosis of the femoral head: a minimum of 10 Years of follow-up.J Arthroplasty. 2020 Mar; 35: 805-810
- Comparison of surgical outcomes between standard and elevated-rim highly cross-linked polyethylene acetabular liners in primary total hip arthroplasty with minimum 15-year follow-up: single-center, retrospective cohort study.J Arthroplasty. 2020 May; 35: 1290-1296
- Minimum 10-year follow-up of cementless total hip arthroplasty with a 32-mm cobalt-chromium head on highly cross-linked polyethylene and a tapered, fiber metal proximally coated femoral stem.J Arthroplasty. 2021 Feb; 36: 647-652
- Complications with hydroxyapatite particulate separation in total hip arthroplasty.Clin Orthop Relat Res. 1994 Jan; : 19-26
- Severe osteolysis after third-body wear due to hydroxyapatite particles from acetabular cup coating.J Bone Joint Surg Br. 1998 Mar; 80: 267-272
- Inferior survival of hydroxyapatite versus titanium-coated cups at 15 years.Clin Orthop Relat Res. 2009 Nov; 467: 2872-2879
- Cementless femoral fixation in total hip arthroplasty.J Bone Joint Surg Am. 2011 Mar 2; 93: 500-509
- Exposure of the hip by anterior osteotomy of the greater trochanter. A modified anterolateral approach.J Bone Joint Surg Br. 1986 May; 68: 382-386
- Results of total hip arthroplasty with bone graft and support ring for protrusio acetabuli in rheumatoid arthritis.J Orthop Sci. 2005; 10: 8-14
- Porous-coated hip replacement. The factors governing bone ingrowth, stress shielding, and clinical results.J Bone Joint Surg Br. 1987 Jan; 69: 45-55
- The case for porous-coated hip implants. The femoral side.Clin Orthop Relat Res. 1990 Dec; : 63-81
- “Modes of failure” of cemented stem-type femoral components: a radiographic analysis of loosening.Clin Orthop Relat Res. 1979 Jun; : 17-27
- Radiological demarcation of cemented sockets in total hip replacement.Clin Orthop Relat Res. 1976 Nov-Dec; : 20-32
- Ectopic ossification following total hip replacement. Incidence and a method of classification.J Bone Joint Surg Am. 1973 Dec; 55: 1629-1632
- Bioinspired surface functionalization of metallic biomaterials.J Mech Behav Biomed Mater. 2018 Jan; 77: 90-105
- Twenty-year follow up of an uncemented stem in primary THA.Clin Orthop Relat Res. 2015 Feb; 473: 543-548
- Effect of stem alignment on long-term outcomes of total hip arthroplasty with cementless Bi-Metric femoral components.J Orthop. 2018 Mar; 15: 134-137
- Results at a minimum of 10 years of follow-up for AMS and PerFix HA-coated cementless total hip arthroplasty: impact of cross-linked polyethylene on implant longevity.J Orthop Sci. 2013 Nov; 18: 962-968
- Strength requirements for internal and external prostheses.J Biomech. 1999 Apr; 32: 381-393
- Difference in postoperative periprosthetic bone mineral density changes between 3 major designs of uncemented stems: a 3-year follow-up study.J Arthroplasty. 2016 Aug; 31: 1836-1841
- Correlation between stress shielding and clinical outcomes after total hip arthroplasty with extensively porous coated stems.J Arthroplasty. 2013 Dec; 28: 1728-1730
- Long-term bone remodelling around ‘legendary’ cementless femoral stems.Effort Open Rev. 2018 Feb; 3: 45-57
- Long-term results of total hip arthroplasty using Charnley Elite-Plus stem and the effect of stem geometry on radiographic distal femoral cortical hypertrophy.J Orthop Sci. 2018 Mar; 23: 365-370
- Risk factors of thigh pain following total hip arthroplasty with short, tapered-wedge stem.Int Orthop. 2020 12; 44: 2553-2558
- Prediction of outcome of total hip arthroplasty based on initial postoperative radiographic analysis. Matched, paired comparisons of failed versus successful femoral components.Clin Orthop Relat Res. 1984 Jun; : 5-15
- Three-dimensional polyethylene wear of a press-fit titanium prosthesis. Factors influencing generation of polyethylene debris.J Arthroplasty. 1997 Apr; 12: 256-266
- Reassessment of computerized wear measurement for total hip arthroplasty with correction for projectional image distortion: a brief follow-up report.J Bone Joint Surg Am. 2010 Aug 4; 92: 1858-1867
Article info
Publication history
Published online: January 17, 2022
Accepted:
December 22,
2021
Received in revised form:
November 30,
2021
Received:
August 17,
2021
Identification
Copyright
© 2022 The Japanese Orthopaedic Association. Published by Elsevier B.V. All rights reserved.